7.15 Avantes Spectrometer
■■■■■■■■■■■■■■■■■■■■■■
1580
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
Analog output
Analog output
on | off (Default value: on)
If this option is enabled, the measured signal on the analog output of the
spectrometer is put out. If the connected spectrometer has no analog out-
put, this option is ignored.
Sensitivity
Only editable when the analog output is switched on.
Range
1 ... 214783647 mAU/mV
Default value
100 mAU/mV
7.15.4
Properties - Detector
Tab: Configuration
▶ Devices ▶ [Edit] ▶ Properties... ▶ Properties - 'Device type' -
'Device name'
▶ Detector
On the tab Detector information on the detector built in the spectrome-
ter is displayed.
Detector type
Display of the detector type.
Minimum wavelength.
Display of the minimum wavelength of the measurable range in nm.
Maximum wavelength
Display of the maximum wavelength of the measurable range in nm.
Number of pixels
Display of the resolution the spectrum is recorded with.
7.15.5
Properties - Calibration
Tab: Configuration
▶ Devices ▶ [Edit] ▶ Properties... ▶ Properties - 'Device type' -
'Device name'
▶ Calibration
On the tab Calibration the data of the wavelength calibration is dis-
played. A specific wavelength, defined by four wavelength coefficients, is
assigned to each pixel on the detector. The wavelength calibration is car-
ried out with the command CAL Spec, the calibration data is calculated
automatically and displayed on this tab.
Date
Display of the time and date of the wavelength calibration.