5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

942

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Measurement with drift control
If the option Measurement with drift control has been selected, then
the following parameters will be displayed:

Signal drift

The measured value is only accepted if the signal drift defined here has
been fallen short of.

Titrando, 855, 867
Range

0.1 ... 999.0 mV/min

Default value

10.0 mV/min

Titrino
Range

0.5 ... 999.0 mV/min

Default value

10.0 mV/min

Min. waiting time

The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The drift contin-
ues to be checked while the waiting time is elapsing.

Titrando, 855, 867
Range

0 ... 999999 s

Default value

0 s

Max. waiting time

If the signal drift has not yet been reached, then the measured value will
be accepted when the maximum waiting time has elapsed. If the waiting
time has not been newly entered then a waiting time that is suitable for
the drift will be calculated automatically according to the following equa-
tion:

Titrando, 855, 867
Range

0 ... 999999 s

Default value

52 s

Titrino
Range

0 ... 9999 s

Default value

52 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.