6 Configuration

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

1333

Note

With Windows 2000, proceed as follows for password monitoring by
Windows:

1.

Run/Execute "secpol.msc".

2.

Under "Local Policies/User Rights Assignment", switch on the
guideline "Act as part of the operating system" for all desired user
groups.

3.

Restart the computer.

Maximum number of login attempts

If this check box is activated, then a user will be set to the status disa-
bled
as soon as the number of login attempts defined here has been
exceeded. A user with the status disabled can no longer start the pro-
gram. The counter for the login attempts is set to zero for all users each
time the program is restarted.

Selection

on | off

Default value

off
FDA default: off (can no longer be edited)

Range

2 ... 5

Default value

3
FDA default: 3 (editable)

Message by e-mail

If this check box is activated, then an E-mail will be sent to the address
defined under [E-mail...] as soon as the defined number of of login
attempts has been exceeded.

Selection

on | off

Default value

off
FDA default: off (editable)

[E-mail...]

This button opens the dialog window Send e-mail (see Chapter 6.2.2.7,
page 1339)
for defining the E-mail parameters.

Logout policies

New login only for same user

If this check box is activated, then only the same user can log in again
after a manual logout. Users with administrator rights can always log in. If
this option is enabled, then the following option is switched on automati-
cally and inactive.

Advertising