3 properties - lift positions, Lift positions – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 1498
6.12 Subwindow Rack data
■■■■■■■■■■■■■■■■■■■■■■
1482
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
is present or not. With off no check will be made. For the option Robotic
arm a swing head with beaker sensor must be installed and a suitable
work position with beaker contact must be defined for the lift, as this
must move to the beaker recognition position.
With the parameter Beaker test in the command MOVE you can deter-
mine whether the determination is to be terminated if a beaker is missing
with or without the display of a corresponding message and whether the
series is to be continued or also terminated.
Selection
Tower | Robotic arm | off
Rack offset
The rack offset is a production-dependent tolerance value between the
upper part of the rack and the lower part. The value is determined by a
rack adjustment and displayed here. If necessary, it can be edited.
Range
-5.00 ... 5.00 °
6.12.3.3
Properties - Lift positions
Tab: Configuration
▶ Rack data ▶ [Edit] ▶ Properties... ▶ Rack data ▶ Lift posi-
tions
Tower 1
Defines the lift positions for Tower 1. These apply for all rack positions
except those that are defined as Special beaker.
Work position
Work position for Lift 1. At this lift position the electrodes, stirrer and
buret tips are optimally positioned for work.
Range
0 ... 235 mm
Rinse position
Rinse position for Lift 1. This lift position is used for rinsing the electrodes.
Range
0 ... 235 mm
Shift position
Shift position for Lift 1. Each time that the rack shifts, the lift will move to
this position if it is located at a lower lift position. If the lift is located at a
higher lift position than that defined here, then the shifting will take place
at the current lift position. This means that the shift position must be
selected so that a safe movement across the entire rack is possible at any
time.
Range
0 ... 235 mm