Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 1078
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1062
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
Parameters
The parameters for the command MEAS TMF are configured on the fol-
lowing tabs:
■
General/Hardware
Parameters for device.
■
Measuring parameters
Parameters for setting the measurement procedure.
Command variables
The following command variables are generated in the method run by the
command MEAS TMF and can be used in formulas under the designa-
tion 'Command name.Variable designation':
Designa-
tion
Description
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
.DBL
Total duration for the processing of the command in s
.EME
End measured value (measured value after processing of
the command) in the unit of the measured value
.ETE
End temperature (temperature following processing of
the command) in °C
.FIN
Command status; 1 = Command has ended at least
once, 0, invalid (variable not available) = Command
has never ended
.IME
Initial measured value (measured value before start con-
ditions are processed) in the unit of the measured value
.ITE
Initial temperature (temperature before start conditions
are processed) in °C
.LP.MEA
Measured value for the last measuring point of the
measuring point list in the unit of the measured value
.LP.TEM
Temperature for the last measuring point of the meas-
uring point list in °C
.LP.TIM
Time in s until the last measuring point in the measuring
point list is reached