6 brc - stop conditions, 7 brc - conditioning, Stop conditions – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 870: Conditioning

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

854

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Extraction time

The titration is not stopped until the extraction time has elapsed (even
when the EP has already been reached). The entry of an extraction time
may be advisable, e.g. for the titration of sparingly soluble samples.

Range

0 ... 999999 s

Default value

0 s

Temperature

Manually entered titration temperature.

Range

-20.0 ... 150.0 °C

Default value

25.0 °C

Time interval measuring point

Time interval for entering a measuring point in the measuring point list.

Range

0.1 ... 999999.0 s

Default value

2.0 s

5.6.3.7.6

BRC - Stop conditions

Tab: Method

BRC Properties... Stop conditions

Command name

Name of the command.

Input

25 characters

The conditions for stopping the titration are defined on this tab.

Stop time

Stop when the entered time has elapsed since the start of the titration.
off means no stop.

Range

1 ... 999999 s

Selection

off

Default value

off

5.6.3.7.7

BRC - Conditioning

Tab: Method

BRC Properties... Conditioning

Command name

Name of the command.

Input

25 characters

Advertising