5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1054

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Parameters
The parameters for the command MEAS Opt Conc are configured on
the following 3 tabs:

(see Chapter 5.6.4.15.2, page 1055)
Parameters for device.

Measuring parameters
Parameters for setting the measurement procedure.

Additional measured values
Defining further measured values which can be saved as additional col-
umns in the measuring point list.

Command variables
The following command variables are generated in the method run by the
command MEAS Opt Conc and can be used in formulas under the des-
ignation 'Command name.Variable designation':

Designa-
tion

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

.DBL

Total duration for the processing of the command in s

.EME

End measured value (measured value after processing of
the command) in the unit of the measured value

.ETE

1)

End temperature (temperature following processing of
the command) in °C

.FIN

Command status; 1 = Command has ended at least
once
, 0, invalid (variable not available) = Command
has never ended

.IME

Initial measured value (measured value before start con-
ditions are processed) in the unit of the measured value

.ITE

1)

Initial temperature (temperature before start conditions
are processed) in °C

.LP.CAx

Calculated value × (1...3) for the last measuring point on
the measuring point list