2 cal loop conc - properties, 7 cal meas conc, 1 cal meas conc - overview – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 1127: Cal meas conc

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

1111

5.6.5.6.2

CAL LOOP Conc - Properties

Dialog window: Method

CAL LOOP Conc Properties.. CAL LOOP Conc -

'Command name'

The number of standards and their concentration are defined in the dialog
window CAL LOOP Conc - 'Command name'.

Command name

Name of the command.

Input

25 characters

Standards

Number of standards

Selection of the number of standards with which the concentration cali-
bration is to be carried out.

Selection

1 | 2 | 3 | 4 | 5

Default value

2

Unit conc.

Selection of the concentration unit for the standards.

Selection

mol/L | mmol/L | µmol/L | g/L | mg/L | µg/L |
mg/mL | µg/mL | ppm | % | mEq/L

Default value

ppm

Conc. standard #

Enter the corresponding concentration for each standard used.

Range

0.001 ... 999999.999

Request for standard exchange

on | off (Default value: on)

If this option is activated, a request for manual solution exchange is dis-
played after each measurement. This option must be disabled for auto-
matic solution exchange when calibrating with a Sample Processor.

5.6.5.7

CAL MEAS Conc

5.6.5.7.1

CAL MEAS Conc - Overview

Dialog window: Method

CAL MEAS Conc Properties.. CAL MEAS Conc -

'Command name'

Measuring command for calibration with ISE electrodes, which is
used together with the command CAL LOOP Conc.

Advertising