Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 819
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
803
The following additional methods for the evaluation of the titration curves
can be activated and defined on these tabs:
Fixed endpoint evaluation
Fix endpoint evaluation
on | off (Default value: off)
If this check box is activated, then the associated values for the other
measured quantities will be interpolated with a fixed measured quantity
(Measured value, Volume or Time) for the fix endpoint from the meas-
uring point list.
Moves the selected line up (changes the sequence).
Moves the selected line down (changes the sequence).
[New]
Open the dialog window Fixed endpoint evaluation # to enter a new
fixed endpoint (see Chapter 5.6.3.5.2.10, page 805).
[Properties]
Open the dialog window Fixed endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.3.5.2.10, page 805).
[Delete]
Delete the selected line.
Minimum evaluation
Minimum evaluation
on | off (Default value: off)
If this option is enabled, then the associated volume, time and tempera-
ture for the minimum measured value will be interpolated from the meas-
uring point list.
Threshold value
The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.
Range
1.0 ... 2000.0 mV/mL
Default value
25.0 mV/mL