9 met ipol - measured value window – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 661
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
645
5.6.3.3.4.9
MET Ipol - Measured value window
Dialog window: Method
▶ MET Ipol ▶ Properties... ▶ Potentiometric evalua-
tion
▶ Evaluation with measured value window (Ipol) ▶ [New]/[Proper-
ties]
▶ Measured value window#
Measured value windows are regions on the measured value axis for
which different parameters for potentiometric evaluation can be defined.
Only those endpoints which lie within these windows and additionally
meet the parameters defined for each window will be recognized. Only
one endpoint will be recognized per window.
Lower limit
Lower limit for the measured value window.
Titrando, 855
Range
-2000.0 ... 2000.0 mV
Default value
-2000.0 mV
Titrino
Range
-2000 ... 2000 mV
Default value
-2000 mV
Upper limit
Upper limit for the measured value window.
Titrando, 855
Range
-2000.0 ... 2000.0 mV
Default value
2000.0 mV
Titrino
Range
-2000 ... 2000 mV
Default value
2000 mV
EP criterion
Criterion for the recognition of endpoints. Endpoints whose discovered
ERC value is smaller than the value entered here will not be recognized.
Range
1 ... 999 mV
Default value
30 mV
EP recognition
Filter for the recognition of equivalence points:
Titrando, 855
Selection
first | greatest | last | ascending | descending
Default value
first