Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 783
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
767
Titrando (without 888)
Selection
1 | 2
Default value
1
855, 888
Selection
1
Default value
1
Sensor
Selection of a sensor of type Metal electrode from the sensors config-
ured in the sensor table. The calibration data for the sensor will be adop-
ted for the determination.
Selection
Sensor name | Metal electrode
Default value
Metal electrode
U(pol)
The polarization voltage is the potential applied to the polarizable elec-
trode during an amperometric measurement.
Titrando, 855
Range
–1250 ... 1250 mV (Increment: 25)
Default value
400 mV
Titrino
Range
–1270 ... 1270 mV (Increment: 10)
Default value
400 mV
Electrode check
on | off (Default value: off)
If this check box is enabled, then an electrode check will be carried out for
polarizable electrodes during the transition from an inactive normal status
to a measurement. A check is also made while doing so to ensure that the
electrode is properly connected and that no short-circuit is present.
Temperature measurement
Type of temperature measurement.
Titrando, 855
Selection
continuous | automatic | off
Default value
automatic
continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.