Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 1142
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1126
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
Moves concentration upward (modifies sequence).
Moves concentration downward (modifies sequence).
[New]
Add the concentration for the next standard.
[Properties]
Edit the concentration of the selected standard.
[Delete]
Delete the selected concentration.
Request for standard exchange
on | off (Default value: on)
If this check box is activated, a request for manual solution exchange is
displayed after each time the calibration loop is run through. This option
must be disabled for automatic solution exchange when calibrating with a
Sample Processor.
Calibration curve
Calibration function
Selection of the function with which the calibration curve is to be calcula-
ted out of the calibration points.
Selection
Linear | Quadratic | Cubic | automatic
Default value
Linear
Confidence interval
Range within which the measured values for the determination of the cali-
bration curve must lie. Values outside this range are called outliers, they
are not integrated in the calculation of the calibration coefficient.
Selection
99 % | 98 % | 95 % | off
Default value
98 %
off
The confidence interval will not be checked.