6 editing rack properties (774), 7 properties - lift positions (774) – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 1502
6.12 Subwindow Rack data
■■■■■■■■■■■■■■■■■■■■■■
1486
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
6.12.3.6
Editing rack properties (774)
Menu item: Configuration
▶ Rack data ▶ Edit ▶ Properties…
Rack name
Shows the name of the rack attached. If no rack is in position then "-----"
will be shown.
Rack code
Shows the number of positions on the rack in position. The rack code cor-
responds to the arrangement of magnets on the base of the rack and is
read in by the Sample Processor in order to recognize the rack. If no rack
is in position then the display will be empty.
Number of positions
Shows the number of positions on the rack. If no rack is in position then
the display will be empty.
The parameters for the attached rack are defined on the following 2 tabs:
■
Lift positions
Defines the work, rinse, shift and special position for Tower 1.
■
Special beakers
Display of the special beaker for the rack in position.
6.12.3.7
Properties - Lift positions (774)
Dialog window: Configuration
▶ Rack data ▶ Properties
Tower 1
Defines the lift positions for Tower 1. These apply for all rack positions
except those that are defined as (see Chapter 6.12.3.8, page 1487).
Work position
Work position for Lift 1. At this lift position the electrodes, stirrer and
buret tips are optimally positioned for work.
Range
0 ... 100 mm
Rinse position
Rinse position for Lift 1. This lift position is used for rinsing the electrodes.
Range
0 ... 100 mm
Shift position
Shift position for Lift 1. Each time that the rack shifts, the lift will move to
this position if it is located at a lower lift position. If the lift is located at a
higher lift position than that defined here, then the shifting will take place
at the current lift position. This means that the shift position must be