5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

859

[Properties]

Open the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 932)
.

[Delete]

Delete the external measured value selected in the table.

5.6.3.7.10

BRC - Fixed endpoint evaluation

Dialog window: Method

BRC Properties... Additional evaluations Fixed

endpoint evaluation

[New]/[Property] Fixed endpoint evaluation #

Measured quantity

Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.

Selection

Measured value | Time | Bromine

Default value

Measured value

Fixed value

Value of the fixed endpoint.

Measured value
Range

-2000.0 ... 2000.0 mV

Time
Range

0.0 ... 999999.9 s

Bromine
Range

0.0 ... 999999.9 µg

5.6.3.8

STAT

5.6.3.8.1

STAT - Overview

Menu item: Method

Insert New command...

Command for STAT Titrations (STAT).

Principle
In STAT titrations, reagent is added until a preset measured value (control
point) has been reached, which is then maintained at a constant level. The
substance released by the reaction is immediately titrated off with the
reagent in order to maintain the control point. The STAT titration keeps
the control point constant until the preset stop criterion has been
reached.