4 cal cond - conductivity standard – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 1118
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1102
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
856
Range
0.1 ... 999999.0 s
Default value
2.0 s
Temperature
Measurement temperature
Temperature of the solution to be measured. If a temperature sensor is
connected, the temperature entered here is ignored.
Range
-20.0 ... 150.0 °C
Default value
20.0
5.6.5.4.4
CAL Cond - Conductivity standard
Tab: Method
▶ CAL Cond ▶ Properties... ▶ Conductivity standard
Command name
Name of the command.
Input
25 characters
Definition of the conductivity standard
Selection of the type as to how the data are to be defined for the conduc-
tivity standard.
Selection
Input | Request | Template
Default value
Input
Input
If this option is selected, then the values for the conductivity standard
can be entered directly.
Request
If this option is selected, then first the current temperature of the solu-
tion is measured after the start of the determination. Afterwards, the
dialog window Data Conductivity standard will appear (see Chap-
ter 5.6.5.4.5, page 1103). The method continues running after the
data has been entered.
Template
If this option is selected, then a conductivity standard can be selected,
and its data will be used to calculate the temperature coefficient for
the specified reference temperature.
Parameters for input
If the option Input has been selected, then the following parameters will
be displayed: