4 set u - control parameters – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 743
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
727
Selection
maximum
Default value
maximum
Titrino
Range
0.01 ... 150.00 mL/min
Selection
maximum
Default value
maximum
Pause 2
Pause 2
Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the addition of the start volume. The pause follows at
the end of all the start conditions.
Range
0 ... 999999 s
Default value
0 s
5.6.3.4.4.4
SET U - Control parameters
Tab: Method
▶ SET U ▶ Properties... ▶ Control parameters
Command name
Name of the command.
Input
25 characters
The control parameters can be set for each endpoint separately.
Endpoint 1
EP1 with
Measured value for the first endpoint.
Titrando, 855
Range
-2000.0 ... 2000.0 mV
Selection
off
Default value
off
Titrino
Range
-2000 ... 2000
Selection
off
Default value
off
Titration rate
Titration rate
The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for Control are not dis-