Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 151
■■■■■■■■■■■■■■■■■■■■■■
3 Workplace
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
135
Sample number
The sample number is set to 0 for all workplaces at the time of each pro-
gram start. Each time a determination is started, it is incremented by +1 .
It can be edited by the user only when it is in the status READY.
Range
0 ... 99999
Default value
0
Remark
Freely definable remarks concerning the determination, that are saved
together with the determination. Prepared sample texts can also be selec-
ted instead of making manual entries (see Chapter 3.5.1, page 130).
Note
The field Remarks can also be modified live while determinations are
running. The context-sensitive menu item Modify remarks is used to
open the dialog window Live modifications (see Chapter 3.6.3.6,
page 138).
Input
100 characters
Autostart
Display of the number of automatic, internal starts of single determina-
tions. This field is only displayed if the option Autorepeat determina-
tions is activated on the Process tab (see Chapter 3.6.3.9.4, page 145).
With every triggering of [Start] the counter is reset to 0.
Statistics
on | off (Default value: on)
This field is displayed only when a method has been loaded in the START
command of which the option Statistics has been activated. The statis-
tics defined in the method can be activated and deactivated manually here
(only in the status READY and COND BUSY).
The first field after Statistics shows the number of determinations already
carried out for the statistics (Actual counter). This field cannot be edited,
but when it is in the status READY, it can be set to 0 with the context-
sensitive menu item Delete statistics, whereby the already calculated
statistics data is deleted at the same time.
The second field after Statistics shows the number of determinations to
be carried out for the statistics (Setpoint counter). In its default settings, it
contains the value Number of single determinations defined in the
START command. This field cannot be edited except in the status
READY and COND BUSY.