Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 516

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

500

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Measuring point density

A small value means small volume increments, i.e. a high measuring point
density. The curve then shows all the finest details which also include
noise; this could cause unwanted equivalence points to be found. A larger
value, i.e. a smaller measuring point density, permits quicker titrations. If
you are using a dosing device with a small cylinder volume then a smaller
measuring point density value may be beneficial. However, you should
also set a smaller signal drift and a higher EP criterion at the same time.

Range

0 ... 9

Default value

2 (slow), 4 (optimal), 6 (fast)

Min. increment

This smallest permitted volume increment is added at the start of the titra-
tion and with steep curves in the region of the equivalence point. Very
small values should only be used if a low titrant consumption is expected;
otherwise unwanted equivalence points could be evaluated.

Range

0.1 ... 999.9 µL

Default value

10.0 (slow), 10.0 (optimal), 30.0 (fast) µL

Max. increment

A maximum volume increment should be selected when the titrant con-
sumption up to the equivalence point is expected to be very small, a start
volume is to be added until just before the equivalence point is reached or
if the change of direction in the potential jump region is very abrupt, as
otherwise it is easy to add too large a volume in the equivalence point
region. The value should not be less than 1/100 cylinder volume.

Titrando, 855
Range

0.1 ... 9999.9 µL

Selection

off

Default value

off

Dosing rate

Speed at which the volume increments are added. The maximum dosing
rate depends on the cylinder volume of the exchange unit or dosing unit
used.

Titrando, 855
Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

Titrino
Range

0.01 ... 150.00 mL/min