Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 56
2.4 Formula editor
■■■■■■■■■■■■■■■■■■■■■■
40
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
Note
In the case of variables with Index {x}, the desired number 1…9 must
be entered for x (e.g. .EP{3}.ERC for the third endpoint).
If no index is specified, then the last index will be used automatically
(e.g. .EP.ERC for the last endpoint).
For the commands ANALOG IN, ANALOG OUT, DIGITAL IN, DIGI-
TAL OUT and STEPPING MOTOR, the syntax is 'Command-
name.Portname.Variabledesignation'.
Designation
Description
Commands
.BLV
Blank value for the sensor used
for the command (only for ISE
sensor) and/or blank value calcula-
ted from the calibration (for CAL
LOOP Conc)
DET U, MET U,
SET U, STAT U,
MEAS U, MEAS
Conc, CAL LOOP
Conc, DOS U
.BP{x}.CNT
Intensity for the break point x
(1…9) in counts
MEAS Opt
.BP{x}.DME
Measured value differential for the
break point x (1…9)
MET
.BP{x}.ERC
ERC or first derivative for the
break point x (1…9)
DET
.BP{x}.MEA
Measured value for the break
point x (1…9) in the unit of the
measured value
DET, MET, MEAS
(without T/Flow,
Ref, Spec)
.BP{x}.TEM
Temperature for the break point x
(1…9) in °C
DET, MET, MEAS
(without T/Flow,
Ref, Spec)
.BP{x}.TIM
Time for the break point x (1…9)
in s
DET, MET, MEAS
(without T/Flow,
Ref, Spec)
.BP{x}.TRN
Transmission for the break point x
(1…9) in %
MEAS Opt
.BP{x}.VOL
Volume for the break point x (1…
9) in mL
DET, MET