9 det ipol - measured value window – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 553

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

537

Additional external measured values

Additional external measured values

on | off (Default value: off)

If this option is enabled, then a maximum of 3 new measured values can
be taken over and saved from existing measuring commands (i.e. MEAS)
which can also be displayed in curves under the designation Extern 1…3
and used in formulas as variables 'Command name.EX1…3'-

[New]

Open the dialog window External measured value # in which the
parameters for the external measured value can be entered.

[Properties]

Open the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 932)
.

[Delete]

Delete the external measured value selected in the table.

5.6.3.2.4.9

DET Ipol - Measured value window

Dialog window: Method

DET Ipol Properties... Potentiometric evalua-

tion

Evaluation with measured value window (Ipol) [New]/[Proper-

ties]

Measured value window

Measured value windows are regions on the measured value axis for
which different parameters for potentiometric evaluation can be defined.
Only those endpoints which lie within these windows and additionally
meet the parameters defined for each window will be recognized. Only
one endpoint will be recognized per window.

Lower limit

Lower limit for the measured value window.

Titrando, 855
Range

-2000.0 ... 2000.0 mV

Default value

-2000.0 mV

Titrino
Range

-2000 ... 2000 mV

Default value

-2000 mV

Upper limit

Upper limit for the measured value window.