3 met ph - start conditions, Start conditions – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 592
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
576
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
Stirrer
Stirrer
Selection of the stirrer.
Titrando, 855
Selection
1 | 2 | 3 | 4 | off
off
No stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Titrando, 855
Range
-15 ... 15
Default value
8
Switch off automatically
on | off (Default value: on)
If this option is enabled, the stirrer will be switched off automatically
when the command has finished. This parameter is displayed only for
Titrandos and 855.
Switch on/off automatically
on | off (Default value: on)
If this option is enabled, the stirrer will be switched on automatically at
the start of the command and will be switched off automatically at the
end of the command. This parameter is displayed only for 751, 785, 798
and 799.
5.6.3.3.2.3
MET pH - Start conditions
Tab: Method
▶ MET pH ▶ Properties... ▶ Start conditions
Command name
Name of the command.
Input
25 characters
The start conditions are processed in the listed sequence before the titra-
tion is started.