3 met ph - start conditions, Start conditions – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 592

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

576

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Stirrer

Stirrer

Selection of the stirrer.

Titrando, 855
Selection

1 | 2 | 3 | 4 | off

off
No stirrer will be used.

Stirring rate

Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.

Titrando, 855
Range

-15 ... 15

Default value

8

Switch off automatically

on | off (Default value: on)

If this option is enabled, the stirrer will be switched off automatically
when the command has finished. This parameter is displayed only for
Titrandos and 855.

Switch on/off automatically

on | off (Default value: on)

If this option is enabled, the stirrer will be switched on automatically at
the start of the command and will be switched off automatically at the
end of the command. This parameter is displayed only for 751, 785, 798
and 799.

5.6.3.3.2.3

MET pH - Start conditions

Tab: Method

MET pH Properties... Start conditions

Command name

Name of the command.

Input

25 characters

The start conditions are processed in the listed sequence before the titra-
tion is started.

Advertising