5 Method
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
683
slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwan-
ted EPs.
optimal
Parameter set for all standard titrations; optimized for the most fre-
quent applications.
fast
For less critical rapid titrations.
User
Editing the individual titration parameters which affect the titration
rate.
Measured value acceptance
Note
Will be displayed only if the option User is selected in the selection list
Titration rate.
Signal drift
Drift for the measured value acceptance during the titration.
Range
0.001 ... 999.000 (mS/cm)/min
Default value
0.05 (slow), 0.1 (optimal), 0.5 (fast) (mS/cm)/
min
Selection
off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed.
Min. waiting time
The minimum waiting time is only important for drift-controlled measure-
ments. The measured value is not accepted until the minimum waiting
time has elapsed, even if the signal drift has already been reached. The
drift continues to be checked while the waiting time is elapsing.
Range
0 ... 999999 s
Default value
0 (slow, optimal, fast) s
Max. waiting time
If the signal drift has been switched off or has not yet been reached, then
the measured value will be accepted when the maximum waiting time has
elapsed. If the waiting time has not been newly entered then a waiting