6 stat ph, 1 stat ph - overview, Stat – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 879

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

863

Cylin-

der vol-

ume

maximum dosing rate

max. flow rate

Exchange

unit

Dosing

unit

Exchange

unit

Dosing

unit

1 mL

2.25 mL/min

---

approx. 130

mL/h

---

2 mL

---

5.00

mL/min

---

approx. 300

mL/h

5 mL

11.25

mL/min

12.50

mL/min

approx. 670

mL/h

approx. 750

mL/h

10 mL

22.50

mL/min

25.00

mL/min

approx. 1.3 L/h

approx. 1.5

L/h

20 mL

45.00

mL/min

50.00

mL/min

approx. 2.7 L/h

approx. 3.0

L/h

50 mL

112.50

mL/min

124.50

mL/min

approx. 6.7 L/h

approx. 7.5

L/h

5.6.3.8.6

STAT pH

5.6.3.8.6.1

STAT pH - Overview

Dialog window: Method

STAT pH Properties... STAT pH - 'Command name'

Command for STAT titrations with measured value pH.

Instruments
This command can be executed with the following devices:

Titrando: 835, 836, 842, 857, 902, 906, 907

Titrino: 718, 736, 751, 799

Robotic Titrosampler: 855

Appearance
The command has the following appearance:

Parameters
The parameters for the command STAT are configured in the following 8
tabs: