6 stat ph, 1 stat ph - overview, Stat – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 879
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
863
Cylin-
der vol-
ume
maximum dosing rate
max. flow rate
Exchange
unit
Dosing
unit
Exchange
unit
Dosing
unit
1 mL
2.25 mL/min
---
approx. 130
mL/h
---
2 mL
---
5.00
mL/min
---
approx. 300
mL/h
5 mL
11.25
mL/min
12.50
mL/min
approx. 670
mL/h
approx. 750
mL/h
10 mL
22.50
mL/min
25.00
mL/min
approx. 1.3 L/h
approx. 1.5
L/h
20 mL
45.00
mL/min
50.00
mL/min
approx. 2.7 L/h
approx. 3.0
L/h
50 mL
112.50
mL/min
124.50
mL/min
approx. 6.7 L/h
approx. 7.5
L/h
5.6.3.8.6
STAT pH
5.6.3.8.6.1
STAT pH - Overview
Dialog window: Method
▶ STAT pH ▶ Properties... ▶ STAT pH - 'Command name'
Command for STAT titrations with measured value pH.
Instruments
This command can be executed with the following devices:
Titrando: 835, 836, 842, 857, 902, 906, 907
Titrino: 718, 736, 751, 799
Robotic Titrosampler: 855
Appearance
The command has the following appearance:
Parameters
The parameters for the command STAT are configured in the following 8
tabs: