Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 1214

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1198

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

not defined
In this case, the device name must be assigned by the user at the start
of the method.

Device type

Display or selection of the device type. If a device is selected under Device
name
, then this Device type field can no longer be edited, and the
device type belonging to the selected device is displayed instead

If the option not defined is selected as Device name, any device types
or device groups which are able to execute the command can be selected,
independently of the devices in the device table.

Selection

Device types | Titrando

Default value

Titrando

Dosing device

Dosing device

Selection of the dosing device to be used for the liquid handling tasks. All
the dosing device connections which are possible with the selected device
type are displayed.

Titrando, 846, 856, 867

Selection

1 | 2 | 3 | 4

Default value

1

778, 789, 814, 815, 855, 864, 874

Selection

1 | 2 | 3

Default value

1

Solution

Input of a solution name or selection of a solution from the solutions
defined in the solution table. If intelligent exchange or dosing units are
used, then a check will be made in the method sequence to see whether
the correct solution has been set on the selected dosing device and
whether the dosing device type is correct. With non-intelligent exchange
or dosing units, only the cylinder volume is checked. At the start of the
command, a check is made of the working life, the validity of the titer and
the GLP test interval for the selected solution.

Input

24 characters

Selection

'Solution name' | not defined

Default value

not defined

not defined
No tests will be carried out.