Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 772
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
756
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
Start drift
The determination can only be started when the currently measured drift
is less than the entered start drift entered here. If this is the case, then
the status displayed in the Run subwindow in the Workplace program
part changes to COND READY.
Titrando, 855
Range
1 ... 999 µL/min
Default value
20 µL/min
Drift correction
If this parameter is set to automatic or manual the drift correction ( =
Drift value × Time for drift correction) caused by empty consumption
during titration is deducted from the total consumption. The Time for
drift correction is the period between end of conditioning and end of
titration.
Selection
automatic | manual | off
Default value
off
automatic
The drift lastly determined during conditioning will be used as drift
value.
manual
The drift value defined under Drift value will be used.
off
No drift correction takes place.
Drift value
Drift value for drift correction. This parameter is only shown for Drift cor-
rection = manual.
Range
0.0 ... 99.9 µL/min
Default value
0.0 µL/min
Stop volume
Stops when the given volume has been added after the start of the titra-
tion. The stop volume should be adapted to the size of the titration cell in
order to prevent any overflow.
Titrando, 855
Range
0.000000 ... 9999.99 mL
Default value
20.00 mL
Selection
off
Default value
off