3 templates, 1 templates - custom calibration buffers, 2 editing custom calibration buffers – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 1370: Templates 4, Custom calibration buffers
6.3 Configuration data
■■■■■■■■■■■■■■■■■■■■■■
1354
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
6.3.3
Templates
6.3.3.1
Templates - Custom calibration buffers
6.3.3.1.1
Managing templates for custom calibration buffers
Dialog window: Configuration
▶ Tools ▶ Templates ▶ Custom calibration buffer
You can define 9 of your own calibration buffers for pH calibration with
automatic buffer recognition in the dialog window Custom calibration
buffers. This buffer series is globally valid and can be selected as buffer
type Custom in the command CAL LOOP pH.
Buffer table
The buffer table shows the defined pH values for the calibration buffers in
5 °C increments. The table can neither be edited nor sorted.
[Edit]
The dialog window Change pH value opens.
[Delete all]
All the pH values of the selected buffer will be set to off.
6.3.3.1.2
Editing custom calibration buffers
Dialog window: Configuration
▶ Tools ▶ Templates ▶ Custom calibration buf-
fer
▶ [Edit] ▶ Change pH value
In the dialog window Change pH values you can edit the pH values for
the buffer selected in the buffer table.
Temperature
Selection of the temperature for which the pH value is to be edited.
Range
0 ... 95 °C
pH value
pH value of the buffer for the selected temperature.
Range
-20.000 ... 20.000
Selection
off
Default value
off