5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

920

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Dosing rate

Rate (volume/time unit) at which the start volume is to be added. The
maximum dosing rate depends on the cylinder volume of the dosing unit
used.

Range

0.100 ... 166.000 mL/min

Selection

maximum

Default value

maximum

Pause

Pause

Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the dosing of a start volume. The pause follows at the
end of all the start conditions.

Range

0 ... 9999 s

Default value

0 s

5.6.3.9.4

TET - Titration parameters

Tab: Method

TET Properties... Titration parameters

Command name

Name of the command.

Input

25 characters

The parameters for the running of the titration are defined on this tab.

Dosing

Dosing rate

Rate (volume/time unit) at which dosing is to be carried out. The maxi-
mum dosing rate depends on the cylinder volume of the dosing unit used.

Range

0.100 ... 166.000 mL/min

Default value

2.000 mL/min

Selection

maximum

Filter

Filtering factor

Factor which determines the digital filtering of the measuring curve. The
higher this value, the more the noise is damped. The filtering has an influ-
ence on the shape of the peak and the EP shift.

Range

1 ... 200

Default value

50