4 stat u - control parameters, Control parameters – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 915

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

899

Pause 2

Pause 2

Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the addition of the start volume. The pause follows at
the end of all the start conditions.

Range

0 ... 999999 s

Default value

0 s

5.6.3.8.7.4

STAT U - Control parameters

Tab: Method

STAT U Properties... Control parameters

Command name

Name of the command.

Input

25 characters

Control point

Control point at

Definition of the current value to be controlled.

Titrando, 855
Range

-2000.0 ... 2000.0 mV

Selection

off

Default value

off

Titrino
Range

-2000 ... 2000 mV

Selection

off

Default value

off

Titration rate

Titration rate

Three predefined rates, for which the parameters for Control are not dis-
played, can selected for the titration rate. The User setting must be selec-
ted in order to be able to edit these parameters.

Selection

50 µL/min | 100 µL/min | 500 µL/min | User

Default value

100 µL/min

Advertising