5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

881

Selection

Cancel determination | Cancel command
(Titrando only)
| Wait for [Continue] | Wait for
limit ok
| none

Default value

none

Cancel determination
The running STAT command will be canceled, then the exit track (if
present) will be started and the determination will be finished.
Cancel command (Titrando only)
The running STAT command will be canceled, then the next com-
mand will be carried out.
Wait for [Continue]
Reagent dosing in the running STAT command will be interrupted and
a message will be displayed. As soon as the monitored measured value
is again within the limits (including hysteresis) reagent dosing can be
resumed by pressing the [Continue] button in this message box.
Wait for limit ok
Reagent dosing in the current STAT command will be interrupted. As
soon as the monitored measured value is again within the limits
(including hysteresis), reagent dosing will be resumed automatically.
none
No action will be taken if limits are exceeded.

Monitoring Dosing rate

Monitoring Dosing rate

on | off (Default value: off)

If this option is activated, then the averaged dosing rate will be monitored
and any exceeded limits will be entered in the measuring point list. The
monitoring of the dosing rate is not started until 10 s after the start of the
titration.

Lower limit

Lower limit of the dosing rate. If the dosing rate falls below this limit, the
event Dosing rate lower limit exceeded is triggered.

Titrando, 855
Range

0.00 ... 166.00 mL/min

Default value

0.00 mL/min

Titrino
Range

0.00 ... 150.00 mL/min

Default value

0.00 mL/min