3 set ipol - start conditions, Start conditions – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 763
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
747
end of the command. This parameter is displayed only for 751, 758, 785,
798 and 799.
5.6.3.4.5.3
SET Ipol - Start conditions
Tab: Method
▶ SET Ipol ▶ Properties... ▶ Start conditions
Command name
Name of the command.
Input
25 characters
The start conditions are processed in the listed sequence before the titra-
tion is started.
Initial measured value
Note
Only displayed for Titrando and 855.
The initial measured value is determined before the other start conditions
are processed. The measured value acceptance depends on the three fol-
lowing parameters:
Measured value drift
The measured value is only accepted if the drift is less than the value
entered here.
Range
0.1 ... 999.0 mV/min
Selection
off
Default value
off
off
The measured vales will not be applied until after the maximum wait-
ing time has passed.
Min. waiting time
The minimum waiting period is only important for drift-controlled meas-
urements. The measured value is not accepted until the minimum waiting
time has elapsed, even if the signal drift has already been reached. The
drift continues to be checked while the waiting time is elapsing.
Range
0 ... 999999 s
Default value
0 s