Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 591
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
575
Sensor
Measuring input
Selection of the measuring input the sensor is connected to.
Titrando (without 888)
Selection
1 | 2
Default value
1
855, 888
Selection
1
Default value
1
Titrino
Selection
1 | 2 | diff.
Default value
1
Sensor
Selection of a sensor of type pH electrode from the list of sensors availa-
ble in the sensor table. The calibration data for the sensor will be adopted
for the determination.
Selection
Sensor name | pH electrode
Default value
pH electrode
Temperature measurement
Type of temperature measurement.
Titrando, 855
Selection
continuous | automatic | off
Default value
automatic
continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise the temperature entered manually
under Temperature on the Titration parameters tab will used.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Titration parameters tab will be
used.