Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 1138

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1122

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Selection

'Device name' | not defined

Default value

not defined

not defined
In this case, the device name must be assigned by the user at the start
of the method.

Device type

Display or selection of the device type. If a device is selected under Device
name
, then this Device type field can no longer be edited, and the
device type belonging to the selected device is displayed instead.

If the option not defined is selected as Device name, any device types
or device groups which are able to execute the command can be selected,
independently of the devices in the device table.

Selection

'Device type' | Titrando

Default value

Titrando

Sensor

Measuring input

Selection of the measuring input the sensor is connected to.

Titrando
Selection

1 | 2

Default value

1

855, 867
Selection

1

Default value

1

Sensor

Selection of a sensor of type pH electrode from the list of sensors availa-
ble in the sensor table. The calibration data for the sensor will be adopted
for the determination.

Selection

'Sensor name' | pH electrode

Default value

pH electrode

Temperature measurement

Type of temperature measurement.

Selection

continuous | automatic | off

Default value

automatic

continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.